Etude sur un dispositif cryochirurgical expérimental.

Study on an experimental cryosurgery apparatus.

Auteurs : ZHANG S. Z., LEI J.

Résumé

The development of a cryosurgery apparatus is important for wider applications of cryosurgery. This paper proposes an apparatus that utilizes auto-cascade refrigeration system. Experimental studies were carried out. Refrigerant mixtures of R-50/R-23/R-600a were used and optimized for the composition so as to achieve lower temperature and higher capacity. By experiment, the temperature at the cryotip could reach as low as -110°C, and 15 W capacity could be obtained at -80°C.

Détails

  • Titre original : Study on an experimental cryosurgery apparatus.
  • Identifiant de la fiche : 2008-2322
  • Langues : Anglais
  • Source : Cryogenics and refrigeration. Proceedings of ICCR'2008.
  • Date d'édition : 05/04/2008

Liens


Voir d'autres communications du même compte rendu (170)
Voir le compte rendu de la conférence