Inventaire d'un système de refroidissement au Ne à écoulement forcé et faible taux de cryogène avec étage de compression à température ambiante et récupération de chaleur.

Low cryogen inventory, forced flow Ne cooling system with room temperature compression stage and heat recuperation.

Auteurs : SHORNIKOV A., KRANTZ C., WOLF A.

Type d'article : Article

Résumé

We present design and commissioning results of a forced flow cooling system utilizing neon at 30 K. The cryogen is pumped through the system by a room-temperature compression stage. To decouple the cold zone from the compression stage a recuperating counterflow tube-in-tube heat exchanger is used. Commissioning demonstrated successful condensation of neon and transfer of up to 30 W cooling power to the load at 30 K using only 30 g of the cryogen circulating in the system at pressures below 170 kPa.