Réfrigérateur à dilution utilisant un tube à pulsation et un cryoréfrigérateur hybride GM pour une expérience de diffusion neutronique.

A dilution refrigerator using the pulse tube and GM hybrid cryocooler for neutron scattering.

Auteurs : KOIKE Y., MORII Y., IGARASHI T., et al.

Type d'article : Article

Résumé

A liquid-He-free dilution refrigerator, "mK Cryocooler", for neutron scattering research is constructed. It obtained the lowest sample temperature of 41 mK in the neutron beam experiment. The mK Cryocooler has a pulse tube cryocooler as the 4 K stage and a Joule-Thomson type dilution refrigerator as the mK stage. The mK Cryocooler is a compact and movable refrigerator for neutron scattering research. The obtained cooling power at 100 mK was about 43 microW when the lowest temperature was 49 mK.

Détails

  • Titre original : A dilution refrigerator using the pulse tube and GM hybrid cryocooler for neutron scattering.
  • Identifiant de la fiche : 2000-2343
  • Langues : Anglais
  • Source : Cryogenics - vol. 39 - n. 7
  • Date d'édition : 07/1999

Liens


Voir d'autres articles du même numéro (1)