Document IIF

Appareil visuel cryogénique pour les équilibres solide-liquide et le mesurage du CH4 + C2H6.

A cryogenic visual apparatus for solid-liquid equilibrium and the measurement of CH4 + C2H6.

Auteurs : WANG X., ZHAO X., YIN W., ZHANG J., ZHAO Y., DONG X., GONG M.

Type d'article : Article de la RIF

Résumé

In this work, an apparatus was designed to measure the phase equilibrium for cryogenic pure fluids and mixtures. The apparatus features view windows and an internal stirrer, and the solid-liquid equilibrium (SLE) is determined through a combination of visual observation and thermal analysis. The apparatus is capable of operating over a range of temperatures (40–300 K) and pressures (0–5 MPa), and has been validated through the measurement of triple points of various fluids. The standard combined uncertainties of temperature, pressure and mole fraction for the SLE measurements are no greater than ±0.70 K, ±0.012 kPa, and 0.010, respectively. The performance of the apparatus is further demonstrated through the presentation of experimental SLE data for methane-ethane binary mixtures.

Documents disponibles

Format PDF

Pages : 110-115

Disponible

  • Prix public

    20 €

  • Prix membre*

    Gratuit

* meilleur tarif applicable selon le type d'adhésion (voir le détail des avantages des adhésions individuelles et collectives)

Détails

  • Titre original : A cryogenic visual apparatus for solid-liquid equilibrium and the measurement of CH4 + C2H6.
  • Identifiant de la fiche : 30031961
  • Langues : Anglais
  • Sujet : Technologie
  • Source : International Journal of Refrigeration - Revue Internationale du Froid - vol. 153
  • Date d'édition : 09/2023
  • DOI : http://dx.doi.org/10.1016/j.ijrefrig.2023.06.005

Liens


Voir d'autres articles du même numéro (32)
Voir la source