OBSERVATIONS SUR LES THERMOMETRES A CMN (NITRATE DE MAGNESIUM CEREUX) ET A RESISTANCE.

SOME OBSERVATIONS ON CMN (CEROUS MAGNESIUM NITRATE) AND RESISTANCE THERMOMETRY.

Auteurs : COOPS G. M.

Résumé

ON REND COMPTE DE RESULTATS EXPERIMENTAUX AU SUJET DE DIVERS TYPES DE MATERIAUX RESISTANTS QUI ONT ETE ESSAYES A BASSE TEMPERATURE. LE COURANT ELECTRIQUE MESURANT LA RESISTANCE PRODUIT DE LA CHALEUR. ON A MESURE L'INFLUENCE DE CETTE ENERGIE THERMIQUE ET DE L'EPAISSEUR DU MATERIAU SUR LE RAPPORTAVEC R-T. LA PUISSANCE PMT DONNANT UNE ELEVATION RELATIVE DE 1 % DE LA TEMPERATURE DU THERMOMETRE S'EST REVELEE NE DEPENDRE QUE LEGEREMENT DES TYPES DE THERMOMETRE, MAIS FORTEMENT DE LA TEMPERATURE. ON DONNE AUSSI LES MESURES ET LES CALCULS QUALITATIFS DE L'INFLUENCE DE LA DESHYDRATATION SUR UN THERMOMETRE A CMN. ON A MESURE LA SENSIBILITE DE CMN DESHYDRATE PARTIELLEMENT (A 50 %) JUSQU'A 15 MK. AU-DESSOUS DE 150 MK ON A MESURE UN ECART AVEC LA LOI DE CURIE. A 29 MK IL APPARAIT UN MAXIMUM DE LA SUSCEPTIBILITE MAGNETIQUE.

Détails

  • Titre original : SOME OBSERVATIONS ON CMN (CEROUS MAGNESIUM NITRATE) AND RESISTANCE THERMOMETRY.
  • Identifiant de la fiche : 1981-1113
  • Langues : Anglais
  • Date d'édition : 03/06/1980
  • Source : Source : Proc. ICEC 8, Genes
    731-735; 3 fig.; 1 tabl.; 8 ref.
  • Document disponible en consultation à la bibliothèque du siège de l'IIF uniquement.