Document IIF

Performance de dégivrage et stratégie de régulation avec bipasse gaz chauds pour un réfrigérateur médical.

Defrosting performance and control strategy with hot gas bypass for a medical refrigerator.

Numéro : pap. n. 285

Auteurs : YUAN L., QIAN S., YU J., et al.

Résumé

Frosting is a common phenomenon in refrigeration and frozen equipment, which degrades heat transfer performance and leads to more energy consumption. This paper investigates the application of hot gas bypass defrosting as an alternative to the electric heater defrosting method. The hot gas from the discharge port enters the evaporator to melt the frost layer on the surface of the evaporator. Experimental data indicate that more than 80% defrosting efficiency can be achieved. The defrosting efficiency increases with thicker frost layer and larger bypassed flow rate. In the end, a defrosting control strategy dedicated to medical refrigerator application is proposed.

Documents disponibles

Format PDF

Pages : 7

Disponible

  • Prix public

    20 €

  • Prix membre*

    Gratuit

* meilleur tarif applicable selon le type d'adhésion (voir le détail des avantages des adhésions individuelles et collectives)

Détails

  • Titre original : Defrosting performance and control strategy with hot gas bypass for a medical refrigerator.
  • Identifiant de la fiche : 30026192
  • Langues : Anglais
  • Source : Proceedings of the 25th IIR International Congress of Refrigeration: Montréal , Canada, August 24-30, 2019.
  • Date d'édition : 24/08/2019
  • DOI : http://dx.doi.org/10.18462/iir.icr.2019.0285

Liens


Voir d'autres communications du même compte rendu (632)
Voir le compte rendu de la conférence