Document IIF

Analyse de l'effet magnéto-calorique grâce au capteur de température optique à infrarouge avec temps de réponse rapide.

Investigation of magnetocaloric effect by the fast response infrared optical temperature probe.

Numéro : pap. 207

Auteurs : KAMANTSEV A. P., KOLEDOV V. V., MASHIROV A. V., et al.

Résumé

We present a new technique for experimental study of kinetics of phase transitions (PTs) and direct measurement of the magnetocaloric effect (MCE) in pulsed magnetic fields by using the fast response temperature probe with infrared fiber optical (IRFO) sensor. As a preliminary demonstration of the new technique, the results are presented of MCE measurements for Gd near Curie point: ?Tad = 21.3 K under pulsed magnetic field µ0H = 12.7 T, with 12.5 ms pulse duration and 1 MHz frequency response of the electronic measurement system.

Documents disponibles

Format PDF

Pages : 4 p.

Disponible

  • Prix public

    20 €

  • Prix membre*

    Gratuit

* meilleur tarif applicable selon le type d'adhésion (voir le détail des avantages des adhésions individuelles et collectives)

Détails

  • Titre original : Investigation of magnetocaloric effect by the fast response infrared optical temperature probe.
  • Identifiant de la fiche : 30019333
  • Langues : Anglais
  • Source : 7th International Conference on Magnetic Refrigeration at Room Temperature (Thermag VII). Proceedings: Turin, Italy, September 11-14, 2016.
  • Date d'édition : 11/09/2016
  • DOI : http://dx.doi.org/10.18462/iir.thermag.2016.0207

Liens


Voir d'autres communications du même compte rendu (71)