Cryostat à hélium liquide pour récepteurs d'IRM utilisant des SQUID.
Liquid helium cryostat for SQUID-based MRI receivers.
Auteurs : SETON H. C., HUTCHISON J. M. S., BUSSELL D. M.
Type d'article : Article
Résumé
The authors describe a liquid helium cryostat, developed to cool SQUID-based receivers in low field MRI systems. The cryostat has a 4 L liquid helium capacity, a hold time of over 3 days and accommodates 10 cm diameter receiver coils. New vacuum insulation methods reduce the noise level by at least an order of magnitude compared to existing commercial designs. Further measurements indicated that most of this field noise probably originates with dielectric losses in the cryostat's fibreglass shells.
Détails
- Titre original : Liquid helium cryostat for SQUID-based MRI receivers.
- Identifiant de la fiche : 2005-2412
- Langues : Anglais
- Source : Cryogenics - vol. 45 - n. 5
- Date d'édition : 05/2005
Liens
Voir d'autres articles du même numéro (3)
Voir la source
Indexation
-
UN CRYOSTAT AVEC HELIUM GAZEUX EN CYCLE FERME C...
- Auteurs : OTOP H., SUJAK B.
- Date : 1984
- Langues : Polonais
- Source : Chlodnictwo - vol. 19 - n. 4
Voir la fiche
-
Three-meter long horizontal cryostat producing ...
- Auteurs : YOSHIKI H., SAKAI K., TAWAI T., GOTO'O S.
- Date : 04/1994
- Langues : Anglais
- Source : Cryogenics - vol. 34 - n. 4
Voir la fiche
-
Novel large volume horizontal cryostat.
- Auteurs : BLAIR D. G., TURNER P. J., DEVLIN J., KANN F. J. van, BRENNAN M. A., SHUSEN C.
- Date : 10/1991
- Langues : Anglais
- Source : Cryogenics - vol. 31 - n. 10
Voir la fiche
-
VACUUM INSULATED SIPHON FOR A CONTINUOUSLY FILL...
- Auteurs : PARPIA J. M., RHODES T. L.
- Date : 1984
- Langues : Anglais
- Source : Rev. sci. Instrum. - vol. 55 - n. 7
Voir la fiche
-
Influence of changes in atmospheric pressure on...
- Auteurs : HANZELKA P., MUSILOVA V.
- Date : 03/1995
- Langues : Anglais
- Source : Cryogenics - vol. 35 - n. 3
Voir la fiche