Description d'un ensemble de bobines de gradient asymétriques associées à un aimant supraconducteur compact pour imagerie à résonance magnétique en neurologie.

Designs for an asymmetric gradient set and a compact superconducting magnet for neural magnetic resonance imaging.

Auteurs : CROZIER S., LUESCHER K., HINDS G., et al.

Type d'article : Article

Résumé

In this article, the authors present the design of a compact, superconducting MRI magnet for head and neck imaging that is less than 0.8 m in length and discuss in detail the design of an asymmetric gradient coil set, tailored to the magnet profile. In particular, the introduction of a radio-frequency FM modulation scheme in concert with a gradient sequence allows the epoch of the linear region of the gradient structure than was previously possible. Images from a prototype gradient set demonstrate the effectiveness of the designs.

Détails

  • Titre original : Designs for an asymmetric gradient set and a compact superconducting magnet for neural magnetic resonance imaging.
  • Identifiant de la fiche : 2000-2463
  • Langues : Anglais
  • Source : Rev. sci. Instrum. - vol. 70 - n. 10
  • Date d'édition : 10/1999
  • Document disponible en consultation à la bibliothèque du siège de l'IIF uniquement.

Liens


Voir d'autres articles du même numéro (2)
Voir la source