Document IIF

Évolution d’un congélateur à ultra-basse température dans le contexte d’une augmentation de la température ambiante.

Ultralow temperature freezer changes with rising ambient temperatures.

Numéro : 1179

Auteurs : SCHUTTE A. J.

Résumé

Ultra-low temperature (ULT) freezers maintain storage temperatures of -50°C to -86°C in ambient conditions ranging from 16°C to 32°C. Freezers with cascade refrigeration systems have been used for decades to preserve cultures, vaccines, samples, specimens, and viruses. Typically, these ULT freezers are stand-alone units, and large institutions operate freezer farms containing hundreds of freezers. Climate change has contributed to an increase in average global temperatures and more frequent hightemperature extremes, such as heatwaves. This paper examined the impact of rising temperatures on ULT freezers. Three key changes were identified: a decrease in efficiency, an increase in heat load, and an increase in component operating temperatures. These changes were investigated using a 400-litre freezer, which maintained a storage temperature of -85°C at ambient temperatures of 20°C and 33°C. The evaluation included the power profile, the pressures for both stages, and the temperature in the compressor compartment.

Documents disponibles

Format PDF

Pages : 12 p.

Disponible

  • Prix public

    20 €

  • Prix membre*

    Gratuit

* meilleur tarif applicable selon le type d'adhésion (voir le détail des avantages des adhésions individuelles et collectives)

Détails

  • Titre original : Ultralow temperature freezer changes with rising ambient temperatures.
  • Identifiant de la fiche : 30034210
  • Langues : Anglais
  • Sujet : Technologie
  • Source : 1st IIR International Conference on Refrigeration Adapting to Rising Temperatures
  • Date d'édition : 08/2025
  • DOI : http://dx.doi.org/10.18462/iir.adaptation.2024.1179

Liens


Voir d'autres communications du même compte rendu (45)
Voir le compte rendu de la conférence