SYSTEME MULTICANAUX BIOMAGNETIQUE AVEC SQUID INTEGRES ET GRADIOMETRES DU PREMIER ORDRE FONCTIONNANT EN ESPACE BLINDE.

BIOMAGNETIC MULTICHANNEL SYSTEM WITH INTEGRATED SQUIDS AND FIRST ORDER GRADIOMETERS OPERATING IN A SHIELDED ROOM.

Auteurs : HOENIG H. E.

Type d'article : Article

Résumé

DESCRIPTION ET PREMIERS RESULTATS OBTENUSAVEC UN DISPOSITIF A 30 CANAUX POUR FAIRE DES DIAGNOSTICS BIOMAGNETIQUES SUR LE CORPS HUMAIN. CE DISPOSITIF COMPORTE UNE SURFACE SENSIBLE REFROIDIE A L'HELIUM LIQUIDE, ENFERMEE DANS UN CRYOSTAT EN FIBRES DE VERRE ET PLACEE DANS UNE CHAMBRE BLINDEE, A COEFFICIENT DE BLINDAGE MOYEN, A L'EXTREMITE D'UN MANIPULATEUR QUI PERMET DE POSITIONNER LA SURFACE SENSIBLE PAR RAPPORT A LA SOURCE BIOMAGNETIQUE. LES SQUID TRAVAILLENT EN MODE BOUCLE DE FLUX FERMEE. ON DISPOSE MAINTENANT D'UN OUTIL POUR ETUDIER LE CERVEAU ET LE COEUR ET FAIRE DES DIAGNOSTICS SUR LE FONCTIONNEMENT DE CES ORGANES.

Détails

  • Titre original : BIOMAGNETIC MULTICHANNEL SYSTEM WITH INTEGRATED SQUIDS AND FIRST ORDER GRADIOMETERS OPERATING IN A SHIELDED ROOM.
  • Identifiant de la fiche : 1990-0966
  • Langues : Anglais
  • Source : Cryogenics - vol. 29 - n. 8
  • Date d'édition : 1989

Liens


Voir d'autres articles du même numéro (7)
Voir la source

Indexation