BIOMAGNETIC MULTICHANNEL SYSTEM WITH INTEGRATED SQUIDS AND FIRST ORDER GRADIOMETERS OPERATING IN A SHIELDED ROOM.

Author(s) : HOENIG H. E.

Type of article: Article

Summary

DESCRIPTION OF AND THE FIRST RESULTS OBTAINED WITH A 30 CHANNEL SYSTEM FOR BIOMAGNETIC DIAGNOSIS OF THE HUMANBODY. THE SYSTEM CONSISTS OF A LIQUID HELIUM COOLED SENSOR ARRAY, ENCLOSED IN A FIBREGLASS CRYOSTAT, LOCATED IN A MAGNETICALLY SHIELDED ROOM WITH A MODERATE SHIELDING FACTOR, SUSPENDED BY A MANIPULATOR ALLOWING THE POSITIONING OF THE SENSOR ARRAY RELATIVE TO THE BIOMAGNETIC SOURCE. THE SQUIDS ARE OPERATED IN THE FLUX LOCKED LOOP MODE. THERE IS NOW A PRACTICAL TOOL FOR RESEARCH AND FUNCTIONAL DIAGNOSIS OF THE BRAIN AND THE HEART.

Details

  • Original title: BIOMAGNETIC MULTICHANNEL SYSTEM WITH INTEGRATED SQUIDS AND FIRST ORDER GRADIOMETERS OPERATING IN A SHIELDED ROOM.
  • Record ID : 1990-0966
  • Languages: English
  • Source: Cryogenics - vol. 29 - n. 8
  • Publication date: 1989