THERMOMETRIE POUR BASSES TEMPERATURES DANS DES CHAMPS MAGNETIQUES ELEVES. RESISTANCES DE CARBONE-VERRE.

LOW-TEMPERATURE THERMOMETRY IN HIGH MAGNETIC FIELDS. CARBON-GLASS RESISTORS.

Auteurs : SAMPLE H. H., BRANDT B. L., RUBIN L. G.

Type d'article : Article

Résumé

ON DECRIT LES CARACTERISTIQUES DES THERMOMETRES A RESISTANCE DE CARBONE-VERRE DE 2,15 A 320 K, DANS DES CHAMPS MAGNETIQUES ATTEIGNANT 19 T. LES RESULTATS DE 9 THERMOMETRES SONT COMBINES POUR PRODUIRE DES VALEURS MOYENNES QUI PEUVENT ETRE UTILISEES POUR CORRIGER UN THERMOMETRE A 19 T A MOINS D'ENVIRON 0,1 % PRES A DES TEMPERATURES DE 30 A 320 K, ET ENVIRON 1 % PRES ENTRE 2,15 ET 30 K. L'ERREUR DE TEMPERATURE CORRESPONDANT A LA MAGNETORESISTANCE D'UN THERMOMETRE A RESISTANCE DE PLATINE DE PRECISION, MINIATURE, A 19 T EST INFERIEURE A CELLE DES ELEMENTS SENSIBLES EN CARBONE-VERRE AU-DESSUS DE 77 K. ON FAIT DE NOUVELLES RECOMMANDATIONS POUR L'UTILISATION OPTIMALE DES THERMOMETRES POUR BASSES TEMPERATURES DANS DES CHAMPS MAGNETIQUES ELEVES.

Détails

  • Titre original : LOW-TEMPERATURE THERMOMETRY IN HIGH MAGNETIC FIELDS. CARBON-GLASS RESISTORS.
  • Identifiant de la fiche : 1983-0431
  • Langues : Anglais
  • Source : Rev. sci. Instrum. - vol. 53 - n. 8
  • Date d'édition : 1982
  • Document disponible en consultation à la bibliothèque du siège de l'IIF uniquement.

Liens


Voir d'autres articles du même numéro (2)
Voir la source