THERMOMETRIE A BASSE TEMPERATURE DANS DES CHAMPS MAGNETIQUES INTENSES. RESISTANCES DE PLATINE DE QUALITE INDUSTRIELLE AU-DESSUS DE 66 K ; RESISTANCES DE RHODIUM-FER ET D'OR-MANGANESE AU-DESSUS DE 40 K.

LOW-TEMPERATURE THERMOMETRY IN HIGH MAGNETIC FIELDS. INDUSTRIAL-GRADE PLATINUM-RESISTORS ABOVE 66 K ; RHODIUM-IRON AND GOLD-MANGANESE RESISTORS ABOVE 40 K.

Auteurs : BRANDT B. L., RUBIN L. G., SAMPLE H. H.

Type d'article : Article

Résumé

ON DECRIT LES CARACTERISTIQUES DE 12 THERMOMETRES A RESISTANCE DE PLATINE DE QUALITE INDUSTRIELLE ET UN DE PRECISION POUR DES TEMPERATURES DE 66 A 300 K DANS DES CHAMPS MAGNETIQUES ATTEIGNANT 19 T. LES RESULTATS DES 13 MODELES PROVENANT DE 9 FOURNISSEURS SONT COMBINES POUR DONNER DES VALEURS MOYENNES POUVANT ETRE UTILISEES POUR CORRIGER DANS L'UN OU L'AUTRE DE CES THERMOMETRES LES EFFETS D'UN CHAMP MAGNETIQUE PARALLELE A SON AXE LONGITUDINAL. L'ERREUR A CORRIGER A 19 T EST INFERIEURE A PLUS OU MOINS 0,3 % ENTRE 66 ET 110 K ET INFERIEURE A PLUS OU MOINS 0,1 % AU-DESSUS DE 110 K. ON PRESENTE AUSSI LA MAGNETORESISTANCE DES THERMOMETRES A RESISTANCE DE RHODIUM-FER ET D'OR-MANGANESE.

Détails

  • Titre original : LOW-TEMPERATURE THERMOMETRY IN HIGH MAGNETIC FIELDS. INDUSTRIAL-GRADE PLATINUM-RESISTORS ABOVE 66 K ; RHODIUM-IRON AND GOLD-MANGANESE RESISTORS ABOVE 40 K.
  • Identifiant de la fiche : 1989-0044
  • Langues : Anglais
  • Source : Rev. sci. Instrum. - vol. 59 - n. 4
  • Date d'édition : 1988
  • Document disponible en consultation à la bibliothèque du siège de l'IIF uniquement.

Liens


Voir d'autres articles du même numéro (3)
Voir la source