COMPARAISON DE QUELQUES THERMOMETRES DE VERRE A BASSE TEMPERATURE DANS UN CHAMP MAGNETIQUE INTENSE.

COMPARISON OF SOME GLASS THERMOMETERS AT LOW TEMPERATURES IN A HIGH MAGNETIC FIELD.

Auteurs : WIEGERS S. A. J.

Type d'article : Article

Résumé

LES AUTEURS ONT COMPARE LA CAPACITANCE A BASSE FREQUENCE DE THERMOMETRES DE VERRE DE FORMES DIVERSES, COMPOSES DE DIOXYDE DE SILICIUM TRES PUR DOPE AVEC ENVIRON 1.200 PPM D'OH-, DE 4 A 100 MILLIK DANS UN CHAMP DE 9 T. LA CONSTRUCTION DES ELECTRODES, LE MODE DE FABRICATION DES CONTACTS ET LE RACCORDEMENT CONDITIONNENT L'INFLUENCE DE LA TENSION ET DU CHAMP SUR LA CAPACITANCE ET LA PERTE. LES AUTEURS ONT CONSTRUIT DES THERMOMETRES DE VERRE CAPACITIFS DONT LA CAPACITANCE N'EST PAS DEPLACEE DANS UN CHAMP DE 9 T A 4 MILLIK, AVEC UNE PRECISION DE LA TEMPERATURE DE PLUS OU MOINS 5 % ENVIRON. CELLE-CI A ETE DETERMINEE EN COMPARAISON D'UN THERMOMETRE A NITRATE DE CERIUM-MAGNESIUM DANS UNE REGION SANS CHAMP ET EN COMPARANT LES MESURES PAR ULTRASONS A L'INTERIEUR ET A L'EXTERIEUR DU CHAMP MAGNETIQUE.

Détails

  • Titre original : COMPARISON OF SOME GLASS THERMOMETERS AT LOW TEMPERATURES IN A HIGH MAGNETIC FIELD.