LOW-TEMPERATURE THERMOMETRY IN HIGH MAGNETIC FIELDS. INDUSTRIAL-GRADE PLATINUM-RESISTORS ABOVE 66 K ; RHODIUM-IRON AND GOLD-MANGANESE RESISTORS ABOVE 40 K.

Author(s) : BRANDT B. L., RUBIN L. G., SAMPLE H. H.

Type of article: Article

Summary

THE CHARACTERISTICS OF 12 INDUSTRIAL-GRADE AND 1 PRECISION PLATINUM RESISTANCE THERMOMETERS FOR TEMPERATURES FROM 66 TO 300 K AND IN MAGNETIC FIELDS UP TO 19 T ARE DESCRIBED. THE RESULTS FROM THE 13 MODELS FROM 9 SUPPLIERS ARE COMBINED TO PRODUCE MEAN VALUES THAT MAY BE USED TO CORRECT ANY OF THOSE THERMOMETERS FOR EFFECTS OF A MAGNETIC FIELD PARALLEL TO ITS LONG AXIS. THE CORRECTABILITY ERROR AT 19 T IS LESS THAN MORE OR LESS 0.3% BETWEEN 66 AND 110 K AND LESS THAN MORE OR LESS 0.1% ABOVE 110 K. MAGNETORESISTANCE DATA FOR RHODIUM-IRON AND GOLD-MANGANESE RESISTANCE THERMOMETERS ARE ALSO PRESENTED.

Details

  • Original title: LOW-TEMPERATURE THERMOMETRY IN HIGH MAGNETIC FIELDS. INDUSTRIAL-GRADE PLATINUM-RESISTORS ABOVE 66 K ; RHODIUM-IRON AND GOLD-MANGANESE RESISTORS ABOVE 40 K.
  • Record ID : 1989-0044
  • Languages: English
  • Source: Rev. sci. Instrum. - vol. 59 - n. 4
  • Publication date: 1988
  • Document available for consultation in the library of the IIR headquarters only.

Links


See other articles in this issue (3)
See the source