IIR document
Device for measurement of thermal emissivity at cryogenic temperatures.
Author(s) : KRÁLIC T., HANZELKA P., MUSILOVÁ V., et al.
Summary
In the described device, the thermal emissivity or absorptivity of the sample is measured by substitution of the radiative heat flow between two parallel surfaces by thermal output of a heater. Fast measurements of the mutual emissivity for the range of the temperature of the radiating surface 25-150 K are possible. The absorbing surface has a temperature between 5 and 10 K when liquid helium is used as cryoliquid. The desired measurement sensitivity is 1 mK for temperature and 0.1 µW for heat power, respectively. The diameter of the whole device is 50 mm and so it is possible to use a commercial dewar can for the cooling. The form of the sample is a round plate 40 mm in diameter and 1 mm in thickness with one tested side. The emissivity and its temperature dependency for various surface treatments can be checked immediately before application in a cryogenic system.
Available documents
Format PDF
Pages: 2004-1
Available
Public price
20 €
Member price*
Free
* Best rate depending on membership category (see the detailed benefits of individual and corporate memberships).
Details
- Original title: Device for measurement of thermal emissivity at cryogenic temperatures.
- Record ID : 2005-1749
- Languages: English
- Source: Cryogenics 2004. Proceedings of the 8th Cryogenics Conference
- Publication date: 2004/04/27
Links
See other articles from the proceedings (27)
See the conference proceedings
Indexing
- Themes: Thermodynamic measurements
- Keywords: Radiation; Cryotemperature; Measurement; Measuring instrument; Emissivity; Cryogenics
-
Emissivity and in situ measurement of the amoun...
- Author(s) : MUSILOVA V., HANZELKA P., KRALIK T.
- Date : 2010/04/26
- Languages : English
- Source: Cryogenics 2010. Proceedings of the 11th IIR International Conference: Bratislava, Slovakia, April 26-29, 2010
- Formats : PDF
View record
-
Emissivity of silver and stainless steel from 8...
- Author(s) : WOODS S. I., JUNG T. M., SEARS D. R., et al.
- Date : 2014/03
- Languages : English
- Source: Cryogenics - vol. 60
View record
-
FAR-INFRARED HIGH-DENSITY POLYETHYLENE WINDOWS ...
- Author(s) : LIM C. C.
- Date : 1986
- Languages : English
- Source: Rev. sci. Instrum. - vol. 57 - n. 7
View record
-
Accurate measuring method for total hemispheric...
- Author(s) : HIGANO M., OZEKI Y., NAKAMURA K., IBE M., MASUDA H.
- Date : 1995
- Languages : Japanese
- Source: Cryogenics/ Cryog. Eng. - vol. 30 - n. 2
View record
-
LOW-TEMPERATURE BOLOMETER ARRAY.
- Author(s) : BONINSEGNI M.
- Date : 1989
- Languages : English
- Source: Rev. sci. Instrum. - vol. 60 - n. 4
View record