ELIMINATION OF FLUX-TRANSFORMER CROSSTALK IN MULTICHANNEL SQUID MAGNETOMETERS.
Author(s) : BRAKE H. J. M. ter
Type of article: Article
Summary
MULTICHANNEL SQUID MAGNETOMETERS ARE BEING DEVELOPED FOR SIGNA L-FIELD MAPPING IN BIOMAGNETIC EXPERIMENTS. A PROBLEM THAT BECOMES MORE SERIOUS AS THE NUMBER OF CHANNELS IS INCREASED IS THE CROSSTALK CAUSED BY THE MUTUAL INDUCTANCES BETWEEN THE INDIVIDUAL SENSING COILS. A METHOD FOR ELIMINATING THIS CROSSTALK IS PRESENTED WHICH IS BASED ON A REARRANGEMENT OF THE FEEDBACK LOOPS WHICH CAUSES THE FLUX-TRANSFORMER CIRCUITS TO BECOME CURRENTLESS. THE FEASIBILITY OF THE METHOD IS VERIFIED EXPERIMENTALLY.
Details
- Original title: ELIMINATION OF FLUX-TRANSFORMER CROSSTALK IN MULTICHANNEL SQUID MAGNETOMETERS.
- Record ID : 1987-1688
- Languages: English
- Source: Cryogenics - vol. 26 - n. 12
- Publication date: 1986
Links
See other articles in this issue (8)
See the source
Indexing
-
Radio-frequency SQUID magnetometer with simplif...
- Author(s) : SIMACEK I., ZRUBEC V., SKAKALA M.
- Date : 1993/08
- Languages : English
- Source: Rev. sci. Instrum. - vol. 64 - n. 8
View record
-
High-critical temperature three-axis direct-cur...
- Author(s) : DANTSKER E., KOELLE D., MIKLICH A. H., NEMETH D. T., LUDWIG F., CLARKE J., LONGO J. T., VINETSKIY V.
- Date : 1994/12
- Languages : English
- Source: Rev. sci. Instrum. - vol. 65 - n. 12
View record
-
BIOMAGNETIC INSTRUMENTATION.
- Author(s) : ROMANI G. L., WILLIAMSON S. J., KAUFMAN L.
- Date : 1982
- Languages : English
- Source: Rev. sci. Instrum. - vol. 53 - n. 12
View record
-
DIGITAL FEEDBACK LOOPS FOR DIRECT CURRENT SQUIDS.
- Author(s) : DRUNG D.
- Date : 1986
- Languages : English
- Source: Cryogenics - vol. 26 - n. 11
View record
-
ELECTRICAL MEASUREMENTS WITH A SQUID SUSCEPTOME...
- Author(s) : WONG H. K.
- Date : 1991/06
- Languages : English
- Source: Rev. sci. Instrum. - vol. 62 - n. 6
View record