Accuracy of commercial dewpoint hygrometers.

Summary

The paper presents an experimental evaluation of different types of commercial chilled-mirror dew-point hygrometers designed to verify their metrological behaviour. Specifically, an analysis of their measurement uncertainties and long-term stability is carried out. A computing method is applied to estimate the measurement uncertainty of an indirect determination of the relative humidity using a chilled-mirror dew-point hygrometer and a dry-bulb temperature sensor.

Details

  • Original title: Accuracy of commercial dewpoint hygrometers.
  • Record ID : 1996-3319
  • Languages: English
  • Source: ASHRAE Transactions.
  • Publication date: 1995/06/24
  • Document available for consultation in the library of the IIR headquarters only.

Links


See other articles from the proceedings (79)
See the conference proceedings