RIPPLE-FREE LIQUID-NITROGEN LEVEL CONTROLLER FOR LONG TERM OPERATION.

Author(s) : SRINIVASAN K., ALTIN J.

Type of article: Article

Summary

THIS ARTICLE DESCRIBES THE DESIGN FEATURES, CONSTRUCTION, AND OPERATION OF A LIQUID-NITROGEN LEVEL CONTROLLER WHICH CAN BE USED CONTINUOUSLY OVER A PERIOD OF SEVERAL WEEKS. LIQUID LEVEL IS SENSED BY PAIRS OF THERMISTORS MOUNTED ON A DIP STICK. REPLENISHING OF LIQUID NITROGEN IS ACHIEVED AS A RESULT OF SELF-PRESSURIZATION OF A TANK CONTAINING THE CRYOGEN. THE TRANSFER SYSTEM ENSURES A RIPPLE-FREE LIQUID SURFACE DURING THE REPLENISHING OPERATION. THIS LEVEL CONTROLLER HAS BEEN A PART OF A CRYOSTAT FOR DIFFUSION MEASUREMENTS IN LIQUEFIED INERT GASES.

Details

  • Original title: RIPPLE-FREE LIQUID-NITROGEN LEVEL CONTROLLER FOR LONG TERM OPERATION.
  • Record ID : 1984-0019
  • Languages: English
  • Source: Rev. sci. Instrum. - vol. 54 - n. 6
  • Publication date: 1983
  • Document available for consultation in the library of the IIR headquarters only.