Superconducting performance test facility for large-current conductor of fusion machine.

[In Japanese. / En japonais.]

Author(s) : TAKAHASHI Y., OSHIRIKI M., KAWANO L., et al.

Type of article: Article

Summary

The superconducting performance test facility for the 50-kiloampere conductor of a fusion machine was completed at the Japan Atomic Energy Research Institute. This facility consists of a 5-kilowatt refrigeration system, 60-kiloampere direct-currect power-supply system, a liquid-helium dewar, and vacuum evacuation and data acquisition systems. the full-size conductor and joint sample for the ITER central solenoid model coils will be tested in this facility. The samples are cooled by a supercritical helium of from 5 to 16 K and tested for an external field of up to 11 T and a 60-kiloampere current.

Details

  • Original title: [In Japanese. / En japonais.]
  • Record ID : 1999-2083
  • Languages: Japanese
  • Source: Cryogenics/ Cryog. Eng. - vol. 33 - n. 5
  • Publication date: 1998/05
  • Document available for consultation in the library of the IIR headquarters only.