COMPARISON OF A GLASS THERMOMETER AGAINST A NUCLEAR ORIENTATION THERMOMETER IN HIGH MAGNETIC FIELDS.

Author(s) : REIJNTJES P. J.

Type of article: Article

Summary

THE AUTHORS COMPARED THE LOW-FREQUENCY CAPACITANCE OF A GLASS THERMOMETER MADE OF VERY PURE SILICA DOPED WITH ABOUT 1,200 PPM OH-WITH A 60 CO.CO SINGLE-CRYSTAL GAMMA-RAY ANISOTROPY PRIMARY THERMOMETER IN THE RANGE OF 15-40 MILLIK IN MAGNETIC FIELDS UP TO 6 T. WITHIN THE TEMPERATURE ACCURACY OF + OR -5% NO SHIFT OF THE CAPACITANCE DUE TO THE APPLIED FIELDS WAS OBSERVED.

Details

  • Original title: COMPARISON OF A GLASS THERMOMETER AGAINST A NUCLEAR ORIENTATION THERMOMETER IN HIGH MAGNETIC FIELDS.
  • Record ID : 1987-0059
  • Languages: English
  • Source: Rev. sci. Instrum. - vol. 57 - n. 7
  • Publication date: 1986