Development of a heat-capacity measurement system under magnetic fields. 1. Magneto-resistance and its anisotropy of a highly stable thin film platinum resistance thermometer, PTF-7.

[In Japanese. / En japonais.]

Author(s) : NARA K., KATO H., OKAJI M.

Type of article: Article

Summary

Measurements were carried out between 20 and 200 K under magnetic fields of up to 8 T. Magnetoresistance showed a high stability during the thermal cycles between room temperature and liquid nitrogen temperature. The variation among the resistance values at 20, 50, 100 cycles was lower than 1 milliK in terms of temperature for 8 of 10 sensors. The anisotropy of the magneto-resistance was also measured and found to be as large as 50% depending on the direction of the field applied to the sensors.