NON-POWER-DISSIPATING CONTINUOUS LEVEL MONITORS FOR LIQUID NITROGEN AND HELIUM.
Author(s) : KUGLER K. J., LINDENBERG W., NIEHUES N.
Type of article: Article
Summary
CAPACITIVE DEVICES WITH ACTIVE LENGTHS UP TO 460 MILLIMETRES HAVE BEEN DEVELOPED FOR THE CONTINUOUS READING OF CRYOGENIC LIQUID LEVELS WITHOUT POWER DISSIPATION. RESOLUTIONS OF + OR -0.7 MILLIMETRE FOR LIQUID NITROGEN AND + OR -1 MILLIMETRE FOR LIQUID AND SUPERFLUID HELIUM HAVE BEEN ACHIEVED. SENSOR CONSTRUCTION, AN ELECTRONIC CIRCUIT FOR CAPACITANCE MEASUREMENT USING THE TECHNIQUE OF CHARGING WITH CONSTANT CURRENT, AND AN AUTOMATIC FILLING UNIT FOR LIQUID NITROGEN ARE PRESENTED.
Details
- Original title: NON-POWER-DISSIPATING CONTINUOUS LEVEL MONITORS FOR LIQUID NITROGEN AND HELIUM.
- Record ID : 1984-1772
- Languages: English
- Source: Cryogenics - vol. 24 - n. 2
- Publication date: 1984
Links
See other articles in this issue (5)
See the source
Indexing
-
DESIGN AND STUDY OF A CAPACITANCE SENSOR FOR ME...
- Author(s) : WU Y. K.
- Date : 1983/03
- Languages : Chinese
- Source: J. chin. Assoc. Refrig. - n. 1
View record
-
Simple device for controlling the level of cryo...
- Author(s) : ROTH P., GMELIN E.
- Date : 1993
- Languages : English
- Source: Cryogenics - vol. 33 - n. 2
View record
-
Two level controller for liquid nitrogen supply...
- Author(s) : YAMAKAWA H., TERASAWA T., SHEN G. H., YUYAMA J.
- Date : 1992
- Languages : Japanese
- Source: Cryogenics/ Cryog. Eng. - vol. 27 - n. 5
View record
-
PASSIVE LEVEL INDICATOR AND VALVE FOR TRANSFERR...
- Author(s) : GOODRICH R. G.
- Date : 1984
- Languages : English
- Source: Rev. sci. Instrum. - vol. 55 - n. 7
View record
-
MAKING OF LIQUID HELIUM TRANSFER TUBE INSTALLED...
- Author(s) : ODASHIMA Y.
- Date : 1983
- Languages : Japanese
- Source: Cryogenics/ Cryog. Eng. - vol. 18 - n. 5
View record