ON THE STABILITY OF CARBON GLASS THERMOMETERS CYCLED AT LOW TEMPERATURES UNDER HIGH MAGNETIC FIELDS.
Author(s) : COUACH M.
Type of article: Article
Summary
REPORT ON THE RELIABILITY OF 3 CARBON GLASS THERMOMETERS USED IN THE TEMPERATURE RANGE 4.2 K-50 K UNDER HIGH MAGNETIC FIELD CONDITIONS. TWO OF THE THREE SENSORS SHOWED A SERIOUS LACK OF REPRODUCIBILITY (1.5 K AT 25 K). THESE RESULTS INDICATE THAT IN EXPERIMENTS REQUIRING PRECISE TEMPERATURE MEASUREMENTS IN THE RANGE 20-30 K, IT IS NECESSARY TO PERFORM REGULAR CALIBRATIONS OF CARBON GLASS THERMOMETERS VERSUS SECONDARY STANDARDS SUCH AS PLATINUM OR GERMANIUM RESISTORS.
Details
- Original title: ON THE STABILITY OF CARBON GLASS THERMOMETERS CYCLED AT LOW TEMPERATURES UNDER HIGH MAGNETIC FIELDS.
- Record ID : 1983-0430
- Languages: English
- Source: Cryogenics - vol. 22 - n. 9
- Publication date: 1982
Links
See other articles in this issue (7)
See the source
Indexing
- Themes: Thermodynamic measurements
- Keywords: Thermometer; Cryotemperature; Cryoprobe; Carbon; Measurement; Glass; Stability; Magnetic field
-
COMPARISON OF SOME GLASS THERMOMETERS AT LOW TE...
- Author(s) : WIEGERS S. A. J.
- Date : 1987
- Languages : English
- Source: Rev. sci. Instrum. - vol. 58 - n. 12
View record
-
BEHAVIOUR OF A CARBON THERMOMETER AND OF A SEMI...
- Author(s) : KHLOPKIN M. N., CHEREMNYKH P. A.
- Date : 1981
- Languages : Russian
- Source: Prib. Teh. eksp. - n. 4
View record
-
A sensitive capacitance thermometer at low temp...
- Author(s) : PENNING F. C., MAIOR M. M., WIEGERS S. A. J., KEMPEN H. van, MAAN J. C.
- Date : 1996/07
- Languages : English
- Source: Rev. sci. Instrum. - vol. 67 - n. 7
View record
-
Possible design for a thin wire resistance ther...
- Author(s) : NARA K., KATO H., OKAJI M.
- Date : 1991/06
- Languages : English
- Source: Cryogenics - vol. 31 - n. 6
View record
-
BEHAVIOUR OF THICK FILM RESISTORS AS LOW TEMPER...
- Author(s) : BOSCH W. A.
- Date : 1986
- Languages : English
- Source: Cryogenics - vol. 26 - n. 1
View record