TESTING OF STRAIN-GAUGE PRESSURE TRANSDUCERS UP TO 3.5 MEGAPASCAL AT CRYOGENIC TEMPERATURES AND IN MAGNETIC FIELDS UP TO 6 T.

Author(s) : CERUTTI G., MAGHENZANI R., MOLINAR G. F.

Type of article: Article

Summary

THE CHARACTERISTICS OF 4 COMMERCIAL STRAIN-GAUGE PRESSURE TRANSDUCERS SUITABLE FOR MEASUREMENTS FROM 4.2 TO 293 K WERE DETERMINED IN THE PRESENCE OF MAGNETIC FIELDS UP TO 6 T. THE HIGHEST PRESSURE WAS 3.5 MEGAPASCAL. TESTS SHOWED THAT IF NO CORRECTIONS ARE MADE, UNCERTAINTIES MAY BE UP TO 19% OF THE FULL SCALE. THE TRANSDUCERS MUST BE INDIVIDUALLY CALIBRATED UNDER THE WORKING CONDITIONS: IN THIS WAY THE UNCERTAINTIES IN PRESSURE MEASUREMENTS CAN USUALLY BE KEPT BELOW 1% OF FULL SCALE.

Details

  • Original title: TESTING OF STRAIN-GAUGE PRESSURE TRANSDUCERS UP TO 3.5 MEGAPASCAL AT CRYOGENIC TEMPERATURES AND IN MAGNETIC FIELDS UP TO 6 T.
  • Record ID : 1984-2229
  • Languages: English
  • Source: Cryogenics - vol. 23 - n. 10
  • Publication date: 1983