THE USE OF INDUSTRIAL-GRADE PLATINUM RESISTANCE THERMOMETERS BETWEEN 77 AND 273 K.

Author(s) : BESLEY L. M., KEMP R. C.

Type of article: Article

Summary

THE STABILITY AND INTERPOLATION CHARACTERISTICS OF SOME INDUSTRIAL-GRADE PLATINUM RESISTANCE THERMOMETERS HAVE BEEN INVESTIGATED BETWEEN 77 AND 273 K. AFTER SOME TENS OF CYCLES BETWEEN 77 AND 273 K THE ICE-POINT RESISTANCES OF THE THERMOMETERS STABILISED TO WITHIN A FEW MILLIOHMS, PROVIDED THAT THE THERMOMETERS WERE NOT HEATED ABOVE 300 K. A TWO-POINT CALIBRATION METHOD IS DESCRIBED, USING RESISTANCE VALUES AT THE ICE-POINT AND THE OXYGEN BOILING POINT.

Details

  • Original title: THE USE OF INDUSTRIAL-GRADE PLATINUM RESISTANCE THERMOMETERS BETWEEN 77 AND 273 K.
  • Record ID : 1983-2104
  • Languages: English
  • Source: Cryogenics - vol. 23 - n. 1
  • Publication date: 1983