Thin film platinum resistance thermometers: calibration and mathematical description of temperature (resistance) function.

Summary

Thin film platinum resistance thermometers with three different dimensions are investigated. Mathematical descriptions of the results in the temperature range 77-300 K are presented. The optimum number of terms is determined for the two types of polynomials which are used for analytical presentation of the experimental data.

Details

  • Original title: Thin film platinum resistance thermometers: calibration and mathematical description of temperature (resistance) function.
  • Record ID : 1995-2831
  • Languages: English
  • Source: Cryogenics - vol. 34 - n. 6
  • Publication date: 1994/06