BEHAVIOUR AND ACCURATE THERMOMETRY OF CARBON-GLASS RESISTANCE THERMOMETERS AT LOW TEMPERATURES AND IN MAGNETIC FIELDS UP TO 7 T.

Author(s) : HUA J., CUI C. G., GUO S. Q.

Type of article: Article

Summary

THE BEHAVIOUR OF CARBON-GLASS RESISTANCE THERMOMETERS OVER A TEMPERATURE RANGE OF 3.41-20 K AND IN MAGNETIC FIELDS OF UP TO 7 T ARE INVESTIGATED. IT IS SHOWN THAT THE PERCENTAGE CHANGES IN RESISTANCE DUE TO THE APPLICATION OF A MAGNETIC FIELD, AS A FUNCTION OF BOTH TEMPERATURE AND FIELD STRENGTH CAN BE DESCRIBED BY AN EQUATION. THE TEMPERATURE ERRORS DUE TO MAGNETORESISTANCE CAN BE CORRECTED TO WITHIN 20 MILLIK.

Details

  • Original title: BEHAVIOUR AND ACCURATE THERMOMETRY OF CARBON-GLASS RESISTANCE THERMOMETERS AT LOW TEMPERATURES AND IN MAGNETIC FIELDS UP TO 7 T.
  • Record ID : 1987-1705
  • Languages: English
  • Source: Cryogenics - vol. 27 - n. 2
  • Publication date: 1987