HELIUM 3 CONSTANT VOLUME GAS THERMOMETER AS INTERPOLATING INSTRUMENT: CALCULATIONS OF THE ACCURACY LIMIT VERSUS TEMPERATURE RANGE AND DESIGN PARAMETERS.

Author(s) : STEUR P. P. M., PAVESE F.

Type of article: Article

Summary

CALCULATIONS ARE PRESENTED FOR A CONSTANT-VOLUME GAS THERMOMETER USED AS AN INTERPOLATING INSTRUMENT BETWEEN FIXED POINTS IN THE RANGE 3.2-25 K AND USING HELIUM 3 AS A WORKING GAS. IN ADDITION TO THE ACCURACY OF THE FIXED POINT REALIZATIONS, IT IS SHOWN TO LIMIT THE ADDITIONAL INTERPOLATION UNCERTAINTY WITHIN 0.1 MILLIK, WITH A SUITABLE CHOICE OF DESIGN PARAMETERS. A STUDY ON THE DEPENDENCE OF THE UNCERTAINTY OF THE MAIN DESIGN PARAMETERS IS ALSO PRESENTED, WHICH SHOWS THAT THEIR CHOICE IS NOT CRITICAL.

Details

  • Original title: HELIUM 3 CONSTANT VOLUME GAS THERMOMETER AS INTERPOLATING INSTRUMENT: CALCULATIONS OF THE ACCURACY LIMIT VERSUS TEMPERATURE RANGE AND DESIGN PARAMETERS.
  • Record ID : 1990-0009
  • Languages: English
  • Source: Cryogenics - vol. 29 - n. 2
  • Publication date: 1989