HELIUM 3 CONSTANT VOLUME GAS THERMOMETER AS INTERPOLATING INSTRUMENT: CALCULATIONS OF THE ACCURACY LIMIT VERSUS TEMPERATURE RANGE AND DESIGN PARAMETERS.
Author(s) : STEUR P. P. M., PAVESE F.
Type of article: Article
Summary
CALCULATIONS ARE PRESENTED FOR A CONSTANT-VOLUME GAS THERMOMETER USED AS AN INTERPOLATING INSTRUMENT BETWEEN FIXED POINTS IN THE RANGE 3.2-25 K AND USING HELIUM 3 AS A WORKING GAS. IN ADDITION TO THE ACCURACY OF THE FIXED POINT REALIZATIONS, IT IS SHOWN TO LIMIT THE ADDITIONAL INTERPOLATION UNCERTAINTY WITHIN 0.1 MILLIK, WITH A SUITABLE CHOICE OF DESIGN PARAMETERS. A STUDY ON THE DEPENDENCE OF THE UNCERTAINTY OF THE MAIN DESIGN PARAMETERS IS ALSO PRESENTED, WHICH SHOWS THAT THEIR CHOICE IS NOT CRITICAL.
Details
- Original title: HELIUM 3 CONSTANT VOLUME GAS THERMOMETER AS INTERPOLATING INSTRUMENT: CALCULATIONS OF THE ACCURACY LIMIT VERSUS TEMPERATURE RANGE AND DESIGN PARAMETERS.
- Record ID : 1990-0009
- Languages: English
- Source: Cryogenics - vol. 29 - n. 2
- Publication date: 1989
Links
See other articles in this issue (8)
See the source
-
DISTORTION OF TEMPERATURE SIGNALS IN HELIUM II ...
- Author(s) : SCHWERDTNER M. V., POPPE W., SCHMIDT D. W.
- Date : 1989
- Languages : English
- Source: Cryogenics - vol. 29 - n. 2
View record
-
Matter and methods at low temperatures.
- Author(s) : POBELL F.
- Date : 1992
- Languages : English
View record
-
HELIUM 3 CRYOSTAT FOR MEASUREMENTS IN SUPERFLUI...
- Author(s) : SCHULZ U., TIDECKS R.
- Date : 1985
- Languages : English
- Source: Cryogenics - vol. 25 - n. 12
View record
-
AN ESTIMATION OF SYSTEMATIC ERRORS IN THE PRESS...
- Author(s) : NARA K., TAMURA O.
- Date : 1985
- Languages : Japanese
- Source: Cryogenics/ Cryog. Eng. - vol. 20 - n. 4
View record
-
Cooled video camera for optical investigations ...
- Author(s) : ALLES H., RUUTU J. P., BABKIN A. V., HAKONEN P. J., MANNINEN A. J., PEKOLA J. P.
- Date : 1994/05
- Languages : English
- Source: Rev. sci. Instrum. - vol. 65 - n. 5
View record