TEST APPARATUS FOR MEASUREMENT OF HEAT CAPACITY OF CRYOGENIC MATERIALS FROM 77 TO 300 K.
Author(s) : PATTANAYAK S., BHOWMICK T.
Type of article: Article
Summary
THE APPARATUS IS A DOUBLE WALLED, HIGHLY EVACUATED, SILVER COATED GLASS DEWAR SUPPORTED IN A RIGID FRAME. THE COVER PLATE MADE OF BRASS HOLDS THE NECK OF THE CRYOSTAT, LIQUID FILL LINE AND VAPOUR OUTLET LINE. THE SAMPLE HOLDER IS MADE OF COPPER FOIL AND THE RATIO OF EMPTY CALORIMETER TO SAMPLE IS MAINTAINED AT 0.11. TEST RESULTS USING THIS APPARATUS HAVE BEEN COMPARED WITH PUBLISHED DATA FOR POLYTETRAFLUOROETHYLENE, POLYETHYLENE AND POLYVINYL CHLORIDE.
Details
- Original title: TEST APPARATUS FOR MEASUREMENT OF HEAT CAPACITY OF CRYOGENIC MATERIALS FROM 77 TO 300 K.
- Record ID : 1991-1932
- Languages: English
- Source: Cryogenics - vol. 30 - n. 2
- Publication date: 1990
Links
See other articles in this issue (5)
See the source
-
AN AUTOMATED ADIABATIC CALORIMETER FOR USE FROM...
- Author(s) : ZHANG Z.
- Date : 1986
- Languages : English
- Source: Sci. sin., B - vol. 29 - n. 12
View record
-
LATTICE THERMAL CONDUCTIVITY OF SEMICRYSTALLINE...
- Author(s) : HASEN A. K., DUBEY K. S.
- Date : 1982/01
- Languages : English
- Source: Indian J. pure appl. Phys. - vol. 20 - n. 1
View record
-
THERMAL CHARACTERISTICS OF POLY(ETHYLENE VINYL ...
- Author(s) : PATTANAYAK S., BHOWMICK T.
- Date : 1990
- Languages : English
- Source: Cryogenics - vol. 30 - n. 9
View record
-
Fast cycle, wide temperature range alternating-...
- Author(s) : FERRARI H., CODNER G., RODRIGUEZ MELGAREJO D., LEVY P., ACHA C., BEKERIS V.
- Date : 1993/11
- Languages : English
- Source: Cryogenics - vol. 33 - n. 11
View record
-
STRUCTURAL AND TECHNICAL CHARACTERISTICS OF CRY...
- Author(s) : GMELIN E.
- Date : 1982/06
- Languages : German
- Source: Kälte + Klimatechnik (Die) - vol. 35 - n. 6
View record