Development of a heat-capacity measurement system under magnetic fields. 2. Performance of a thermal-relaxation calorimetric system under magnetic fields.

[In Japanese. / En japonais.]

Author(s) : KATO H., NARA K., OKAJI M.

Type of article: Article

Summary

The system is based on the thermal relaxation method by use of a temperature-difference detection system, which incorporates a pair of small thin film resistance thermometers, and is little affected by magnetic fields. Its performance under magnetic fields has been studied by use of a reference sample, whose heat capacity is independent of magnetic fields. The performance of the total system has been studied between 40 and 90 K under magnetic fields of up to 7 T.