FAST CERIUM MAGNESIUM NITRATE (CMN) THERMOMETER FOR THE LOW MILLIKELVIN TEMPERATURE RANGE.

Author(s) : GREYWALL D. S., BUSCH P. A.

Type of article: Article

Summary

THE CONSTRUCTION DETAILS AND THE PERFORMANCE OF A CMN SUSCEPTIBILITY THERMOMETER OF NOVEL DESIGN ARE DISCUSSED. THE DEVICE FEATURES A RESPONSE TIME OF 30 SECONDS AT 3 MILLIK AND LESS THAN 2 SECONDS AT 10 MILLIK. WHEN OPERATED AS PART OF A SIMPLE SELF-INDUCTANCE BRIDGE CIRCUIT AT A DRIVE LEVEL SUFFICIENT TO OBTAIN A RESOLUTION OF ONE PART IN 10,000, THE POWER DISSIPATION IS ONLY 10 PICOWATTS AND THE SELF-HEATING IS NEGLIGIBLY SMALL. THIS NEW THERMOMETER MAKESPOSSIBLE A NEW CLASS OF EXPERIMENTS IN THE LOW MILLIK REGIME.

Details

  • Original title: FAST CERIUM MAGNESIUM NITRATE (CMN) THERMOMETER FOR THE LOW MILLIKELVIN TEMPERATURE RANGE.
  • Record ID : 1990-0008
  • Languages: English
  • Source: Rev. sci. Instrum. - vol. 60 - n. 3
  • Publication date: 1989
  • Document available for consultation in the library of the IIR headquarters only.