Platinum resistor thermometry and pressure calibration in a clamped pressure cell with the medium, Daphne 7373.

Author(s) : MURATA K., YOSHINO H., YADAV H. O., HONDA Y., SHIRAKAWA N.

Type of article: Article

Summary

Calibration of a platinum resistance thermometer (Netsushin) in magnetic fields and under pressure in the range of 1.5-300 K and below 1.5 gigapascal is presented. With the pressure medium, olefin olygomers, Daphne 7373, the pressure is continuous at its solidification and the pressure drop from 300 to 4.2 K by 0.15-0.17 gigapascal is constant, irrespective of the initial clamped pressure at 300 K. The applicability of the thermometer and the medium for precise study in field and pressure at low temperature is discussed.